   Calibrators for automated metrological verification of electricity meters (UAPS)

UAPS-2

PURPOSE AND FIELD OF APPLICATION

    The calibrator is intended for manual or automated periodic recalibration and adjustment 
of single-phase electricity meters of the class  1 and less accurate (including those that have affiliated serial and parallel circuits) measuring active electric energy of the forward and reverse direction. Calibrators for metrological verification can be used as high-technology metrology device when calibrating and adjusting single-phase shunt energy meters.

    UAPS-2 is a standard wattmeter-counter of the class 0,2 and fictitious power source  which are structurally of unitized design. When binding several units it is possible to connect up to 36 electricity meters at the same time.

TECHNICAL FEATURES OF CALIBRATORS FOR METROLOGICAL VERIFICATION OF ELECTRICITY METERS

    UAPS-2 ensures:
  • reduction of calibration time  by half as a result of automatic setting of calibration and adjusting modes of electricity meters  owing to  control from a personal computer  (PC);
  • meter error measurement, comparison of error value with preset permissible deviations, as well as indication for rejection;
  • calibration of meter sensitivity and meter creep;
  • keeping of a calibration protocoll  in form of a saved file and the possibility to print it out;
  • absence of additional errors because of exclusion of human factor during calibration.

COMPLETE EQUIPMENT OF CALIBRATORS FOR METROLOGICAL VERIFICATION OF ELECTRICITY METERS

  • UAPS-2 calibrator
UAPS-2 calibrator can be equipped with multipurpose meter calibration bench for 12 meters, 2 groups of 6 electricity meters for operative connection of meters calibrated.
  • Programme control of calibrator and automated meter calibration.

TECHNICAL DATA

Quantity

Value

Class of meters calibrated

1,0

Number of energy meters calibrated at the same time

6

Nominal phase voltages, V

230

Working range of phase voltages

from  0,8 Unom t 1,2 Unom

Main working range of currents,

from 0,01 t 100

Additional current range,

from 0,001 t 0,01

Maximum permissible relative error when setting voltage and current  values within working range, %

±1,0

Maximum permissible relative error when setting current  value within additional range, %

±5,0

Maximum voltage signal power at each output, V

10

Maximum power of current signal, W

170

Rated frequency of current and voltage signal, Hz

50 or 60

Range of measurement of phase shift angle between voltage and current signals, degrees.

±180

Error when setting phase shift angle, degrees

±1

Maximum permissible relative error when measuring active power and active energy of the forward and reverse direction at power factor from 0,5 t 1, 0, %

from 0,2 t 0,3

Maximum permissible relative error when measuring  voltage and current, %

±0,3

Maximum permissible percentage error when measuring current within additional range, %

±0,3

Average error-free running time, h

not less than 5000

Average service life, years

not less than 6

Measuremetns of the calibrator, mm

470496160,5

Mass of the calibrator, kg

not more than 22,0

PRINCIPLE OF OPERATION OF CALIBRATORS FOR METROLOGICAL VERIFICATION OF ELECTRICITY METERS

  • UAPS-1 calibrator for metrological verification of electricity meters operates under control of IBM PC personal computer (processor frequency 300 Hz, RAM 64 b, support for RS-232 (COM-port), video display supporting 800600 resolution, OC Win 98/Me).
  • Data transmission between the computer and calibrator is carried out via serial digital RS-232 communication interface.
  • The calibrator is powered from single-phase 220V AC network.
  • Power consumption – not more than 900 V
 